Determination of chloride ions in compound fertilizers by fully automatic potentiometric titration


1 Preface

Compound fertilizers are chemical fertilizers made by chemical methods and contain any two or three of the crop nutrients nitrogen, phosphorus and potassium.

Its function is to meet the comprehensive needs and balance of various nutrients required by agriculture under different production conditions. During the production process of compound fertilizers, manufacturers, considering cost issues, may use raw materials such as ammonium chloride and potassium chloride. Chlorine is a trace nutrient element. If the content of chloride ions is too high, it will affect seed germination and cause soil compaction. Therefore, the content of chloride ions in compound fertilizers is an important indicator for measuring the quality of compound fertilizers. This paper adopts the fully automatic potentiometric titration method to detect the chloride ion content in compound fertilizers. The operation is simple, the results are accurate and the precision is high.

2. Instruments and reagents

2.1 Instruments

JH-T7 fully automatic potentiometric titrator, silver composite electrode, 40-mesh sieve

2.2 Reagents

Silver nitrate standard solution (0.02mol/L), nitric acid solution (30%)

3 Experimental Methods

3.1 Experimental Procedures

Grind the fertilizer into fine powder with a mortar, pass it through a 40-mesh sieve, accurately weigh 2.5g of the fertilizer into a 250mL beaker, add 100mL of water, heat slowly to boiling, continue to boil gently for 10 minutes, cool to room temperature, transfer the solution to a 250mL volumetric flask, and dilute to the mark. Mix well, dry filter, and discard the initial part of the filtrate. Take 1mL of the filtrate into a titration cup, add 60mL of water and 2mL of nitrate solution, and titrate to the endpoint with 0.02mol/L silver nitrate titrant. At the same time, conduct a blank test.


3.2 Parameter Settings

Titration mode Dynamic titration

 stirring speed :

5

Electrode equilibrium time:

6s

Pre-stirring time :

6s

Electrode equilibrium potentia:

1mv

Titration speed :

标准

Minimum addition volume:

0.04mL

Pre-titration addition volume :

6mL

Final volume:

10mL

Pre-titration stirring time:

30s

Potential spike

150

Pre-controlled mv value:

no

 

4. Results and Discussion

4.1 Experimental Results

Sample Name  Titrant Concentration (mol/L) Sample volume (mL) Blank Volume (mL)  Titration volume (mL)  Chloride ion (%)

 

Average value (%)

 

 

Compound fertilizer

 

 

0.020

 

 

2.56009

 

 

0.360

9.320

62.04

 

 

62.27

9.368

62.40

9.368

62.36

 

4.2 Calculation Formula:

佳航

X- The content of chloride ions in the sample, expressed as a percentage (%);

V1- Volume of silver nitrate standard titration solution consumed in titrating the sample dilution solution, measured in milliliters (mL); V0- Volume of silver nitrate standard titration solution consumed blank, unit: milliliters (mL);


C - Concentration of silver nitrate standard titration solution, unit: moles per liter (mol/L);

4.3 Map

佳航

 

 

4.4 Conclusion

The chloride ion content in compound fertilizers was determined by fully automatic potentiometric titration. The absolute difference in parallel tests of the same sample was less than 0.4%, meeting the requirements of the national standard. The instrument can automatically control the titration process, determine the endpoint and process data, reducing the error caused by naked-eye determination of the endpoint. It features rapidity and simplicity. It can also reduce the contact between personnel and organic reagents, enhancing personnel safety.