The content of chloride ions in the solution of electronic components is determined by potentiometric titration


I. Preface

Chlorides act as anode activators in the electroplating solution of electronic components. Therefore, the detection of chloride ions is an important test for electroplating solutions of electronic components

Measure indicators. This method uses silver nitrate as the titrant and takes advantage of the reaction principle that chloride ions and silver ions react to form silver nitrate, with silver ions being used for selection

The electrode was used as the measuring electrode, and the chloride ion content was detected by the JH-T6 fully automatic potentiometric titrator. The detection results were highly accurate and the detection efficiency was fast.

It is convenient to operate, reduces human interference at the same time, and enhances the stability of the detection.

Ii. Instruments and Reagents

2.1 Instruments

JH-T6 fully automatic potentiometric titrator, silver ion selective electrode, analytical balance, etc

2.2. Reagents

Nitric acid (analytical grade), silver nitrate standard solution (0.1mol/L), purified water.

Iii. Experimental Methods

3.1 Analysis Steps

Accurately measure 1.0mL of the sample to be tested with a 1mL pipette into a titration cup, add 50mL of water, place it on the titration stand, and start the pre-set

The method is to titrate with the calibrated silver nitrate solution until the potential suddenly crosses the endpoint, and record the amount used. At the same time, conduct blank experiments.

The parameters of the titrator are set as shown in Table 1

Table 1 Titration Parameter Settings

Titration type:  Dynamic titration  method name:  Chloride content determination
Burette volume:

10mL

Sample measurement unit:

g

Working electrode:  Composite silver electrode Reference electrode

无

Stirring speed:

8

Pre-stirring time

5s

Display unit: 

mV

 End volume:

20mL

Electrode equilibrium time: 

4s

 electrode equilibrium potential:

1mv

Titration speed:  standard titration  Equilibrium potential

6mv

Pre-titration addition volume:

0

Minimum addition volume:

0.02mL

Potential spike: 

500

 pre-controlled mv value:

无

Correlation coefficient: 

35.5

Result unit:

g/L

Titrant name: 

Silver nitrate solution  Theoretical concentration: 0.1(calibrated concentration)

3. Test graph example

佳航

Iv. Results and Discussion

4.1 Experimental Results

The experimental results are shown in Table 2

Table 2 Test Results

Sample name

Sampling volume

mL

c(AgNO3)

mol/L

Blank volume

V0 /mL

Titration volume

V1 /mL

Chloride

g/L

Average content

g/L

 

 

Solution

 

 

1

 

 

0.1027

 

 

0.02

2.796

10.121

 

 

10.442

2.866

10.376

2.990

10.828

 

4.2. Conclusion

The determination of chloride ions by potentiometric titration is rapid and has good data repeatability. It reduces the interference of indicator color on the results and improves the detection

Efficiency.

References

[1] Wan Zhuo. Detection Methods for Nickel Ions, Chloride Ions and Boric acid Content in Electroplating Nickel Solution [J]. Industrial Technological Innovation.

 

Keyword:

potentiometric titration