17

2025

-

06

Determination of sodium chloride content in proteins by fully automatic potentiometric titration


1. Preface

Sodium chloride is indispensable for the human body. It is of great significance for maintaining normal osmotic pressure of cells, regulating the body's acid-base balance, and preserving the physiological activity of proteins.

This method, in accordance with the provisions of the 3107 Sodium Chloride Determination Method in the 2020 Edition of the Chinese Pharmacopoeia, removes the protein from the sample, completely reacts the chloride ions in the solution with an excess of silver nitrate, and titrates the excess silver nitrate with ammonium thiocyanate titrant. After adopting the JH-T7 fully automatic potentiometric titration method of Shanghai Jiahang, it is more accurate, efficient and safe than manual titration, and the data repeatability is good, meeting the detection requirements in the pharmacopoeia.

2 Instruments and reagents

2.1 Instruments

JH-T7 fully automatic potentiometric titrator, silver composite electrode, 10mL burette

2.2 Reagents

Ammonium thiocyanate titrant (0.05mol/L), silver nitrate solution (0.1mol/L), nitric acid solution (8.0mol/L).

3 Experimental Methods

3.1 Experimental Procedures

Accurately measure 1.0mL of the test sample, precisely add 5mL of 0.1mol/L silver nitrate solution, mix well, add 10mL of 8.0mol/L nitric acid solution, heat and digest until clear, cool, and titrate with ammonium thiocyanate titrant (0.05mol/L) until the potential jump endpoint. Correct the titration result with a blank. The titration result was corrected with a blank.


3.2 Instrument Parameters

佳航

4 Results and Discussion

4.1 Experimental Data

4.1.1 Blank Calibration

Sample name titration solution concentrated sample volume  average titration volume Average value

 

Silver nitrate

 

0.05

 

5

11.083

 

11.098

11.113

 

4.1.2 Determination of Aluminum solution

Sample Name titration solution concentrated sample volume  titration volume Sodium chloride content Average value

 

Sodium chloride sample dissolved

 

liquid

 

 

0.05

 

 

1

7.596

10.2346

 

 

10.2317

7.511

10.4830

7.684

9.9774

 

4.2 Calculation Formula

Sodium chloride content (g/L) = (V0- V X) × c × 58.45, where V0 is the volume of ammonium thiocyanate titrant consumed in the blank test, mL;

VX- The volume of ammonium thiocyanate titrant consumed by the sample, in mL;

C- Concentration of ammonium thiocyanate titrant, mol/L;

V2- The volume of the sample, mL;

58.45 is the molecular weight of sodium chloride.

4.3 Titration Spectrum

4.3.1 Blank

佳航

4.3.2 Sodium chloride solution sample

佳航

4.4 Conclusion

The determination of sodium chloride content in protein-containing samples by JH-T7 fully automatic potentiometric titrator has accurate results, good data repeatability, is faster and more convenient, and meets the detection requirements.

Keyword:

automatic potentiometric titration